Astra C8-HE

Programmable Mass Flow Controller with Digital Signal Processing

Microprocessor driven digital flow controllers allow one to program, record, and analyze flow rates of various gases with a computer via an RS-485 interface (optional RS-232 is available).

Principles of Operation

Metered gases are divided into two laminar flow paths, one through the primary flow conduit, and the other through a capillary sensor tube. Both flow conduits are designed to ensure laminar flows and therefore the ratio of their flow rates is constant. Two precision temperature sensing windings on the sensor tube are heated, and when flow takes place, gas carries heat from the upstream to the downstream windings. The resultant temperature differential is proportional to the change in resistance of the sensor windings. A Wheatstone bridge design is used to monitor the temperature dependent resistance gradient on the sensor windings which is linearly proportional to the instantaneous rate of flow. The output of the Wheatstone bridge is converted to digital format with a 12 Bit ADC (analog to digital converter).

Design Features

  • Digital and Analog modes operate simultaneously.
  • Programmable Flow Configurations.
  • Multi-Drop Capability of up to 256 units.
  • Stores calibration data for up to 10 gases.
  • Totalizer indicates total gas quantity.
  • Alarm limits for high and low gas flow.
  • Conversion factors for up to 256 gases.
  • Auto Tune function for optimum control response.
  • Self Diagnostic Tests.

Technical specifications

Accuracy15°C to 25°C and 0.7 to 4 bars: ±1% of FS, 0°C to 50°C and 0.3 to 10 bars: ±2% of FS,±1% of FS at a specific temperature and pressure with special calibration
Repeatability±0.15% of full scale
Response Time0.6 to 1.0 s to within ±2% of set point over 20% to 100% of FS
Temperature Coeffcient0.05% of FS/°C
Pressure Coefficient0.01% of full scale/0.07 bar
Optimum Gas Pressure1.73 bars
Maximum Gas Pressure70 bars
Maximum Differencial Pressure3.4 bars for DFC2600 and DFC3600, 2.8 bars for DFC4600
Gas/Ambient Temperature5°C to 50°C
Communication InterfaceRS485 - Standard, RS232 - Optional
Output SignalLinear 0-5 V= (2000 ohms min load impedance), 0-10V= (4000 ohms min impedance), 4-20 mA optional (0-500 ohms loop resistance)
Circuit ProtectionBuilt-in polarity reversal protection. Resettable fuses provide power input protection.
Materials in Fluid Contact316/416 stainless steel, Viton/Neoprene/Kalrez O-rings (depends on model)
Attitude SensitivityNo greater than +15 degree rotation from horizontal to vertical (standard calibration is in horizontal position)
ConnectionsModel DFC26 standard 1/4” compression fittings, Model DFC36 standard 1/4” compression fittings, Model DFC4600 standard 3/8” compression fittings, Optional 1/8” or 3/8” compression fittings and 1/4” VCR7 fittings available.
Leak Integrity1x 10-9 ml/s of helium
Transducer Input Power+15V=, 450 mA maximum
Calibration OptionsStandard 10 point NIST calibration (Optional up to 9 additional 10 point calibrations may be ordered for an additional charge)
CE ComplianceEN 55011 class 1, class B; EN50082-1
LC Phases C8-HE
Particle Type Fully Porous

Brochures

Other Documents

Product name ID Particle Size Pore Size LC length LC Technique Preparative column type Product code
ASTRA® C8-HE HPLC column, 3,0 µm 150 mm × 4,6 mm 4.6 mm 3 µm 100 A 150 mm HPLC AST-5896-IK46
ASTRA® C8-HE HPLC column, 5,0 µm 150 mm × 4,6 mm 4.6 mm 5 µm 100 A 150 mm HPLC AST-5896-LK46
ASTRA® C8-HE HPLC column, 5,0 µm 250 mm × 4,0 mm 4 mm 5 µm 100 A 250 mm HPLC AST-5896-LM40
ASTRA® C8-HE HPLC column, 5,0 µm 250 mm × 4,6 mm 4.6 mm 5 µm 100 A 250 mm HPLC AST-5896-LM46
ASTRA® C8-HE HPLC column, 5.0 µm 100 mm × 30 mm, DAC column HW 30 mm 5 µm 100 mm DAC ASP-5896-LI3X-LDAC
ASTRA® C8-HE HPLC column, 5.0 µm 250 mm × 50 mm, DAC column HW 50 mm 5 µm 250 mm DAC ASP-5896-LM5X-LDAC